CLX-XT 28030 datasheet: what an engineering purchase specification needs
Updated
The CLX-XT 28030 is an extended-reagent-life total chlorine and TRO monitor. Its datasheet is a purchasing starting point; the matching manual, installed revision and vessel conditions determine the final specification.

Key takeaways
- Record range and accuracy together.
- Check sample pressure, distance and drain.
- Confirm complete power demand and included consumables.
Specify the range with its accuracy
The supplied XT manual lists a standard 0–10 mg/L range and optional 0–15 mg/L extended range. For standard operation, accuracy is ±5% of reading or ±0.03 mg/L, whichever is greater, through 6.0 mg/L, and ±10% above that. The extended-range specification is ±10% or ±0.03 mg/L, whichever is greater. State the required range and applicable accuracy together; a display range alone is not an acceptance criterion.
Check sample delivery before ordering
The manual specifies approximately 0.34–10.3 bar inlet pressure and 200–400 mL/min flow to waste. It calls for installation close to the sampling point, around 2–3 metres, and a drain open to atmosphere. The instrument cannot compensate for an unrepresentative sample or a blocked drain. Record sample-point distance, available pressure and the approved waste return arrangement in the installation scope.
Use the complete electrical and enclosure information
The XT specifications table lists 100–240 VAC, 47–63 Hz and 250 VA, while a wiring subsection describes individual supply requirements differently. Confirm the complete installed assembly and nameplate rather than sizing a circuit from an isolated paragraph. The manual gives an IP66 enclosure with an IP55 fan and does not recommend outdoor use. A high enclosure rating is not permission to install it unprotected on deck.
State what comes with the monitor
The engineering specification identifies 28030, tubing/cuvette kit 09950XT, spare flow-through cuvette 25018S and reagent supply 09991. These references assist identification, not universal substitution. Confirm current part availability, serial compatibility and whether reagents and service consumables are included. Download the specification and manual below and retain the applicable editions with the purchase order.
Illustrative procurement scenario
A vessel has sufficient sample pressure but a long, shared drain. Rather than purchasing the monitor alone, the RFQ includes sample take-off and drain review so that an installation limitation is not mistaken for a defective analyser.
Request a matched CLX quotation
Sea Clean is a Watts distributor. We source, supply and repair all CLX variants and supply normal, zig-zag and J.A.W. TRO reagent kits competitively. Stocked products are ready for delivery; confirm the exact item and quantity. Estimated DHL Express transit is 1–4 days on suitable routes, subject to dispatch, customs and final vessel handover.
For this enquiry, include: Existing serial number or 28030 requirement; measurement range; power; sample pressure and temperature; installation location; required reagent and service kits.
Equipment enquiries: CLX monitor supply
Consumables: TRO reagent kits
Service support: BWTS service and calibration
Frequently asked questions
Does IP66 mean the CLX-XT is an outdoor instrument?
No. The supplied manual does not recommend outdoor use and separately rates the fan. Review the complete installation and applicable revision.
Manufacturer PDFs & direct enquiries
Original manufacturer documents hosted directly by Sea Clean. Match the model and revision before use.
Email CLX enquiry to post@seaclean.no
Opens an email draft. Attach nameplate and bottle photographs, add your requirements, then send. No website form.
Explore all 15 CLX buying and service guides
- HF scientific CLX family: select the model and generation before buying
- CLX-XT 28030 datasheet: what an engineering purchase specification needs
- OEM-labelled CLX and MU units: identifying the monitor behind the BWMS label
- CLX-Ex and Ex2 supply and repair: verify the hazardous-area assembly
- CLX-XT versus XT2: enclosure and retrofit checks before replacement
- Techcross zig-zag TRO reagent kits: match the installed bottle arrangement
- J.A.W. powder reagent kits: distilled-water preparation and purchase checks
- Compare normal, zig-zag and J.A.W. TRO kits on delivered usable cost
- HF scientific CLX spare parts: avoid a near-match service kit
- CLX stock and DHL Express delivery: plan the final handover to the vessel
- CLX-XT repair or replacement: make the decision from the fault evidence
- CLX-XT cooling chamber: investigate shortened reagent life safely
- Legacy CLX to current CLX: USB-C, Modbus and analogue-output upgrade checks
- CLX DPD accuracy: turn a datasheet tolerance into an acceptance requirement
- Watts distributor CLX support: one enquiry for supply, repair and service
Sources
- CLX-XT engineering specification — ES-HF-CLX-XT; ordering information and operating limits. Manufacturer PDF hosted by Sea Clean; reviewed 9 November 2026.
- CLX-XT installation, operation and maintenance manual — Watts-hosted 28652 revision 5 file; match the installed serial number and software. Manufacturer PDF hosted by Sea Clean; reviewed 9 November 2026.
Related articles
- OEM-labelled CLX and MU units: identifying the monitor behind the BWMS label
- CLX-Ex and Ex2 supply and repair: verify the hazardous-area assembly
- CLX-XT versus XT2: enclosure and retrofit checks before replacement
- Techcross zig-zag TRO reagent kits: match the installed bottle arrangement
- J.A.W. powder reagent kits: distilled-water preparation and purchase checks
- Compare normal, zig-zag and J.A.W. TRO kits on delivered usable cost